In detail: DIY repair of a tonometer cuff from a real master for the site my.housecope.com.
I took out the cuff. Poisons air in a rubber bag near the place where the hose enters. Is it possible to glue it somehow (if possible, how and how?) Or can you just throw it away and buy a new one?
You can also try. But it won't last long. The cuff changes its volume when inflating, and the glue will soon crumble.
Now I specially called a medical technician friend, and he told me what he was doing at home in such cases.
When measuring blood pressure with a mechanical tonometer, we can sometimes suddenly notice that yesterday this device quickly and easily reached the required numbers and normal filling of the cuff when pumping up. And today - the brush gets tired of idle pumping, and the arrow actively rushes to zero. This means there is a malfunction in the apparatus. Do this:
- First, exclude the possibility that the rubber hose connecting the cuff and the bulb has jumped off the cuff or starts to jump off. Air may leak in this area. Push the hose deeper, if so. If air continues to pass through there, try tightly wrapping this section with a double thick thread, tying it up more often, and seal it with tape on top.
- If it is found that the air flows directly from the cuff, then determine this place better, take a strip of thick aluminum, bend it in half, push the hole into the aluminum and clamp it with metal, like pliers. This will be a kind of patch. From above, you can spread it with glue.
Maybe this cuff will last for a while. But it is obvious that soon you will have to buy a new one. The cuff is essentially a thick rubber balloon. Such products are very difficult to glue.
Video (click to play). |
This video was created for informational purposes only. From practice I add that the cuffs are in the tonometer.
repair of the cuff of an automatic tonometer.
How and how to glue a chamber for a tonometer.
Tonometer. The pressure does not rise when pumped or drops rapidly when measured without a skip. Checked the connections.
Repair of the cuff by gluing.
Budget repair of the tonometer cuff.
how I made the cuffs from a motorcycle tube.
Replacing the camera in the cuff of the shoulder tonometer.
One of the subscribers recommended a sealant to glue the door rubber in the washing machine. So I spent it.
Cuffs for the tonometer Passive earnings, 11 months working without problems, I checked it myself, maybe someone is interested.
Repair of the AND UB-201 device if the cuff is torn. Error: ERR_CUF.
Where to repair, how and why to glue the tonometer cushion, how to properly glue the tonome rubber cushion.
Repairing a manual tonometer, as it turned out, is not so difficult if there are Chinese spare parts on sale from Aliexpress.
In most cases, the repair of the tonometer is the replacement of a failed spare part: a pear.
Elimination of the hole in the chamber of the tonometer cuff.
Cleaning the bleed valve.
Buy an inexpensive tonometer: https://ali.pub/295awr Repair of the OMRON electronic automatic tonometer and adapter. On the.
Choosing a tonometer begins with choosing a cuff. OMRON cuffs guarantee accurate and painless measurement.
Tonometer Citizen CH 605 gives an error when measuring. This is due to the hole in the chamber.
Repair (cleaning) of the air relief valve.
The AEG BMG 4922 blood pressure monitor does not inflate the cuffs. Troubleshooting. - Repair of the cuff.
The period in the operation of the air relief valve.
Tonometer A&D Medical UB-402. If there is a problem with the pump, I recommend replacing it, and do not suffer from repairs))).
The tonometer gives an error due to an unsealed chamber.
The tonometer does not turn on. RADIO CHANNEL FOR BEGINNERS is free.
LIFEHACK)) I don't remember who and when taught me this .. it was a long time ago .. by the way, it turned out to be a little higher for me.
How to remove the cuff from the tonometer and how to properly disassemble and assemble the case in reverse order.
Measuring chamber repair.
In a neighborly way, they gave me a look at the broken Microlife blood pressure monitor / The L-shaped connector was broken. This video is fast.
TM OMRON cuffs for blood pressure monitors on the shoulder are absolutely unique. The soft fan-shaped cuff allows for repetition.
Repairing a broken cuff connector.
It is no secret that blood pressure is an important indicator for the functioning of the heart, blood vessels, the entire body system and human vitality. There is always at least one person in the family suffering from hypertension. Therefore, the house should have a tonometer for regular monitoring of the main indicator of health. Especially the device is required for the elderly and those who are engaged in heavy physical labor. As a rule, devices are prone to breakdowns, especially mechanical types of tonometers break down. Therefore, the problem of how to repair a tonometer with your own hands is quite relevant.
As a rule, the repair of devices consists in changing the component unit. Why you need to buy a new mechanism and replace the broken one with it. For example, it is not possible to inflate the cuff with air. This means that the pumping check valve in the bag has broken. It is pointless to change only the valve, you need to replace the entire assembly, along with the tubes.
Electronic and semi-automatic machines are several times higher than a mechanical device. For these types of device, not too frequent malfunctions are noted. Often happen:
- malfunctions in the mechanism itself;
- absence or delay in displaying indicators on the display;
Display problems
It is better to eliminate serious breakdowns in a service center, because a warranty card is issued for repairs, and before its expiration, it is prohibited to carry out repairs on your own. You can change the bulb, cuff, batteries yourself, they are not covered by the warranty card.
If you decide to save time and money and fix the tonometer yourself, then you must first find the malfunction in the device. If a person uses it often, then it will not be difficult to identify the broken part. If the problem is air leaking from the system, then it must be detected. Having found a place for the passage of air, you just need to replace this element with a new one. If after that the pressure measurement went well, then everything worked out.
What breakdowns can a mechanical device have and how to fix them yourself:
-
The cause of the breakdown lies in the falling out of the gauge needle. Here you cannot fix the breakdown on your own. It is recommended to buy a new gauge and replace the broken one with it.
Pressure gauge in a mechanical tonometer
Sticking valve in the pear
You can buy the necessary part for a mechanical blood pressure monitor in an online store by ordering by phone, or in a medical equipment store.
Breakdowns and repair methods for an electronic type device:
-
When unrealistic indicators are displayed on the board, it is required first of all to replace the battery. If this does not help, you need to look for air leaks in the cuff, pear and the pipe connecting them. Here, repairs are carried out according to the circumstances: replacement with new parts or elimination of a leak.
A quick solution to the problem - changing the battery
It is not so easy to disable automatic devices. If the tonometer is not dropped, struck or shaken with force, then it will serve for many years. The electronics inside the camera are quite fragile and require delicate and careful handling.
If the problem is bleeding air, then you need to find the place of the gap. When it is located directly on the cuff, it is not necessary to replace it, it can be repaired. How to repair a vented tonometer cuff:
- If the hole is small and the cuff is strong and not worn, then it can be sealed with a hot melt gun.
- You can glue a piece of rubber patch or other suitable material onto the gap with rubber glue.
- Take a strip of thick aluminum and fold it in half. Put the damaged area between them and clamp it with metal, help yourself with pliers. Cover this place with glue.
When the air escapes at the factory soldering site, it can be vulcanized with a soldering iron at 180 degrees. It is important to remember that repairing the cuff is cheaper than replacing it, but it will not last too long after the intervention. Therefore, be prepared to replace this element.
Tonometer model AND UB 201
The AND UB 201 tonometer is a bracelet-type device for measuring pressure indicators. Most often, this model fails the cuff. To repair or replace it, you must remove it correctly. This is done in sequence:
- The batteries are removed from the camera.
- Remove the clamps that secure the cuff (they are located on both sides of the case) by prying them off with a screwdriver.
- Pulling the cuff towards you, release it from the tonometer body, to which it is attached with two more latches.
- The housing cover is attached with two screws at the back, they need to be unscrewed with a screwdriver. Open the cover carefully and remove it.
- To completely disassemble the tonometer, press in the recesses at the top of the case with the same screwdriver, and it will easily fall apart.
Disassembling the device case
The body is assembled in the reverse order. It is important to put all its parts in place and fix them on special "pins".
The question is considered in detail, if the tonometer does not work, what to do. You can contact the service center and hand it over for repair, but it will take a lot of time, extra money will be spent. The breakdown associated with the replacement of an element can be eliminated by yourself - you can buy a new part in a medical equipment store and put it in place of the faulty one.
Can the tonometer cuff be glued on?
In theory, of course you can. Practically the same. It largely depends on how old your cuff is and whether it has worn out physically at all. Since rubber ages over time and loses its elastic properties. If the hole is near the hose itself, this will be especially difficult to do. And in general, if quot; houndsquot; in one place, then there is a reason to be wary. Better to buy a new rubber bulb and insert it into the tonometer. This device is very important at the RIGHT moment. Therefore, it is not worth saving on a rubber pear.
Soldered with a hot glue gun. While holding.
You can also try. But it won't last long. The cuff changes its volume when inflating, and the glue will soon crumble.
Now I specially called a medical technician friend, and he told me what he was doing at home in such cases.
When measuring blood pressure with a mechanical tonometer, we can sometimes suddenly notice that yesterday this apparatus quickly and easily reached the required numbers and normal filling of the cuff when pumping up. And today - the brush gets tired of idle pumping, and the arrow is actively moving towards zero. This means there is a malfunction in the apparatus. Do this:
- First, exclude the possibility that the rubber hose connecting the cuff and the bulb has jumped off the cuff or starts to jump off. Air may leak in this area. Push the hose deeper, if so. If air continues to pass through there, try tightly wrapping this section with a double thick thread, tying it up more often, and seal it with duct tape on top.
- If it is found that air flows directly from the cuff, then determine this place better, take a strip of thick aluminum, bend it in half, push the hole into the aluminum and clamp it with metal, like pliers. This will be a kind of patch. You can spread it on top with glue.
Maybe this cuff will last for a while. But it is obvious that soon you will have to buy a new one. The cuff is essentially a thick rubber balloon. Such products are very difficult to glue.
Kaneshno it is possible, I myself not very long ago had such a problem on an automatic blood pressure monitor, so I pasted it there and just in case reeled up with electrical tape.And nothing, it has been working fine for several months now.
Etching at the place of factory soldering. Vulcanized with a soldering iron at a temperature of 180 degrees. While it works. Maybe someone will come in handy
Dear author. When you checked your blood pressure, you didn’t put the cuff on your upper arm correctly. It's a pity.
Pzh answer I was played disconnected then connected and then it did not work was the first problem from the video how to fix
but if the machine shows the upper pressure more than it is. checked with a doctor.
Tell me what kind of rectangle appeared at the bottom of the screen, before it was not there, and the air is pumped into nothingness. I have a semiautomatic device Rosmax.
Thank you very much for the useful information!
I changed the camera in the cuff. The old one was made of silicone, it seemed to be soft and very elastic. And the new one was made of polymer, a little rougher. Will the meter readings change? Is this adjustable or not?
This is not a renovation, but a banal replacement of the old with a new one. I myself found the air bleed point in the cuff, by lowering the pumped cuff into a bath of water and sealed it with a small patch and glue. Everything is OK.
Dear friend! So they repair it in a good workshop, where there are spare cuffs and it's good if their price is 200 rubles. As a rule, any man will perform these simple operations. But for example, Automatic OMROT, the upper pressure shows normal, and the lower one underestimates somewhere by 10 units. So tell me what and where to tweak, what I was able to measure correctly! Advice, buy a new one, does not work. So what to do? This is checked in comparison with other blood pressure monitors, preferably manual ones. Because machines often have errors
If you are so smart, tell me, people will say thank you! Sorry in advance if offended !! PS: And the video is really useful, thanks to the author.
please tell me I can't find anything like this, the problem: tr “” Microlife BP A50 ″ ”pumps normally bleeds like the same speed, but with the indication that something is wrong during the bleed, the indication reaches about 140 -150 - then the signal pee-pee and all the indication is gone! and the bleeding of air continues to go on as usual to the end and that's it.
Thanks for the help, the semiautomatic device of the Andes, etched the cuff, changed it, but still the venting step was 5-7 units, took out the check valve from the pear - everything was purely only a small hair sticking out in the slots, pulled it out and inserted the valve back. After the repair, the venting step became 3 units - as it was. Thanks again!
Everything is shown perfectly. I have a similar device, a semiautomatic device, AED, when you pump up to 200 normally, but from above you can say it does not pump, or rather pumped up, but it immediately lowers. It's more of a mechanical problem than an electronics problem, right? If possible, tell me.
Check the cuff and check valve as in the video
what a clever craftsman, he has spare pears and cuffs at home. which no one else has at home.
The rubber tubes (inner diameter) of the cuffs from different manufacturers are unified. For example, the Omron tonometer cuff can be used with the Microlife tonometer.
Video uploaded by KALYANYCH
Search KALYANYCH 8 months ago
How and what to glue a chamber for a tonometer
how i made cuffs from a motorcycle camera
How to quickly reduce blood pressure at home - with and without medication. Treatment of hypertension and reduction of high blood pressure without pills. High blood pressure - what to do. Medicines that reduce high blood pressure. Pressure pills. What pressure is normal and what is considered high. High blood pressure. How to treat hypertension. *** ATTENTION! This video is for educational and informational purposes only. There are contraindications. Before using the recommendations and tips from the video, ALWAYS consult with your doctor! *** Subscribe to Dr. Evdokimenko's channel / REASONABLE MEDICINE to keep abreast of the news:
The Gamma m1-1 tonometer does not gain pressure due to a non-working relief valve.
Dear author. When you checked your blood pressure, you didn’t put the cuff on your upper arm correctly. It's a pity.
Pzh answer I was played disconnected then connected and then it did not work was the first problem from the video how to fix
but if the machine shows the upper pressure more than it is. checked with a doctor.
Tell me what kind of rectangle appeared at the bottom of the screen, before it was not there, and the air is pumped into nothingness. I have a semiautomatic device Rosmax.
Thank you very much for the useful information!
I changed the camera in the cuff. The old one was made of silicone, it seemed to be soft and very elastic. And the new one was made of polymer, a little rougher. Will the meter readings change? Is this adjustable or not?
This is not a renovation, but a banal replacement of the old with a new one. I myself found the air bleed point in the cuff, by lowering the pumped cuff into a bath of water and sealed it with a small patch and glue. Everything is OK.
Dear friend! So they repair it in a good workshop, where there are spare cuffs and it's good if their price is 200 rubles. As a rule, any man will perform these simple operations. But for example, Automatic OMROT, the upper pressure shows normal, and the lower one underestimates somewhere by 10 units. So tell me what and where to tweak, what I was able to measure correctly! Advice, buy a new one, does not work. So what to do? This is checked in comparison with other blood pressure monitors, preferably manual ones. Because machines often have errors
If you are so smart, tell me, people will say thank you! Sorry in advance if offended !! PS: And the video is really useful, thanks to the author.
please tell me I can't find anything like this, the problem: tr “” Microlife BP A50 ″ ”pumps normally bleeds like the same speed, but with the indication that something is wrong during the bleed, the indication reaches about 140 -150 - then the signal pee-pee and all the indication is gone! and the bleeding of air continues to go on as usual to the end and that's it.
Thanks for the help, the semiautomatic device of the Andes, etched the cuff, changed it, but still the venting step was 5-7 units, took out the check valve from the pear - everything was purely only a small hair sticking out in the slots, pulled it out and inserted the valve back. After the repair, the venting step became 3 units - as it was. Thanks again!
Everything is shown perfectly. I have a similar device, a semiautomatic device, AED, when you pump up to 200 normally, but from above you can say it does not pump, or rather pumped up, but it immediately lowers. It's more of a mechanical problem than an electronics problem, right? If possible, tell me.
Check the cuff and check valve as in the video
what a clever craftsman, he has spare pears and cuffs at home. which no one else has at home.
The rubber tubes (inner diameter) of the cuffs from different manufacturers are unified. For example, the Omron tonometer cuff can be used with the Microlife tonometer.
DIY tonometer cuff repair.
today we have a repair of a rubber cuff for a tonometer, in general, large 1 time, I broke one more thing when merrell pressure in general, well, they read it, in general, and I had problems here in this part, but only in another
me well, as it were, original and Indian andy and an andean manufacturer to a more or less high quality, but still it got covered so in the fact that there I was covered here and to fix it somewhere in the middle of here
you see how it is done so it is more or less visible at least his ataman as divine I will take it and show it better, this is some kind of Chinese noun-name of a horse for Avito served for a year, maybe even less, I already took it soaked you know for sure that he is in this place with me
keep the focus as it is done here such as a hole for this Velcro and somewhere from this here from this place the cloaks go well, to get there I need to somehow pair this whole thing here and watch this is a drawback, but that's all
I also want to do it this way, but here it is much easier to see something like that in quality and ours are more or less here, but here it is easier for me to do it, because I know for sure that I have been gurgling for years, but here in such places I will also take a watch I will inflate him
I watched how the service center did a bottle of water like this, you put it on and blow, you look, you let it pass, and then I already let it go completely, but now there was nothing left of it, but now I think I can just take it at the match, and like that
so in order not to completely torment and the first thing I thought with ordinary molecular glue to take the night, this option is fig, well, if I know how I think of a cyclist, take it with ordinary glue and repair cameras and take it and maybe
a piece of this bike trays and we will try to repair it by the way, this elastic band is all so tightly dressed, removal is generally impossible, but I twisted the Chinese noun-name here like this for this
I don’t know, well, in short, the past does not take this one that is already ready and pull there for some reason to shoot and pull on this pimp here, if it’s visible there goes to dice such a plastic thing, it’s like this is such a width and it is glued to the fabric as
I suspect that just one thing the matter goes between the matter, he even can see this hole, you can and more than once steamed this thing here pour glue and try to glue it like this always try badly just that it is all wet this thing let it be
the hole will dry up, then I know it’s already not here, now we’ll be like this and looking at the sofa and the hole is like this and it is removed this is the whole thing, too tight, then here it pulls on the topic, so the tour is really tight, this is all very important here
not to break the pastor's wife with them, it’s better to shoot in this way, because it’s much easier to remove it like this, but, on the other hand, it’s many times easier to dress just like cheeks all this I have a tonometer andy
her 105 model has been working on the problem for a long time, only something is constantly with these not pears ready the gun I was stunned so I changed the pear even though it works for a smart manufacturer all this here is the cuff the cuff is dead, but the pears are still working
One more day I’m blowing up this whole thing, I’m looking at it barely, but it’s still going down, and now we’ll try by sound somewhere she’s in this place in this fold somewhere she’s here
misses, but how to understand and look the holes is not particularly visible and how this camera works, too, I have not yet understood it as a fabric, but in the middle there is a kind of rubber spraying, probably on this fabric, and somewhere if it breaks through, it can come out
to be somewhere along the seam, so I also think for myself how to glue it, then if I even find a hole like this on this material, you see it like that, well, like matter, they only knew approximately then
here is the part of it, but I’m like this, here’s a lot of emotions, it seems like the material goes here, but you see this camera is already visible, that is, the camera itself in the middle has torn and so that we don’t see each other, it’s generally unreal, then all this is needed here
I argued and only there and it can be glued, that is, this is how you can rip it into such sacrifices, even though it is worth doing so at night, I took a camera in the middle, after all, there is such a camera in the material and with
the sides where this metal thing ripped apart, in general, this is the structure I have from Andy, but the micro bubble seems to differ a little this is the main one of China, this is what is interesting on youtube and wherever I looked, they say that it is not
glued, he tried to glue the moment naturally, as if you see how, as it were, for me, but this is where someone soldered this with an iron or even with a soldering iron in general such nonsense, but lost weight
I signed up with the service center that bought this cuff last year, I say I need to remove the glue globe, I thought it was like a moment or not the same moment or not, well, I also have an inter-globe, but this will make some other composition for bicycle cameras here in general
nothing of the kind is written very simply to apply wait it seems to me if the same manufacturer is probably the same I will also try to look for a globe here now I will not run, but you know the globe it is in gray in
gray such a tube of sulfur is the same as it is only gray it is written here in large letters a globe without trays without anything advised me just need to apply several layers a little different here in india there is such a bubble here video in itself it seems to be too
made by the pimples of the mesh, as it were, the principle is the same here, but this is how it is done and I have a suspicion of this case because the gurgles somewhere are coming from here with a glance so I walked around this place is possible, but hardly because it is sealed here
here is such a thing, I'm here some numbers and it's not clear to the factory numbers, but I know for sure that I had a problem in this place, look where it was bent, but if you take it like this, even noticeably holes, your daughter looks and in general we will glue it in in general, it is so small
small at all, this is she alone, well, there are no such suspicious places, except that I didn’t like it.
me a gurgle yesterday here in this place, which is interesting no trays to use nothing said this service Maine says it can only aggravate the situation maybe this is true, I ask there will be a tray, it seems as if our very thin expands on
together the foot itself will appear, there may be cracks and well, what’s bad, it’s here in my place, just here, on the bend, and it will remain on the bend, how can I glue it on a clear stump like this
but then she will still return to this position, no, I don’t even know what to say, here is such a nasty hole, so it’s even rubbing would say it’s just rubbed off now I need to apply several layers and so first I’ll apply one layer just the wash will freeze him 2
upstairs let's try it here it smells at the box office at all that moment is absolutely January 1 so I'll take it a little well it goes that he doesn't want to go about it went like this and the camera glue the camera on the bike that it will be and then still goes
oh there, sometimes there are such divorces as if home a lot is not needed here well, and now, as it were, take a layer to apply to form, as it were, I don’t know how it will be glued at all and whether it will be because there is such a structure here is it to clean it up or not to protect it
I decided to make a bicycle tire straight like this, even at least the width should have been tried, so I will give such a droplet like this in a circular motion, how to formulate the layer is really bad
the dream turns out at us just at the junction, as it were, it might turn out a shura really diana then a hole like this I don’t know this is all good here such a layer turned out a hole is not visible in that case here I’ll put it so it is present after 10 hours 2 again on
I've been waving about two hours, this is the structure of the Maz the second time, which is interesting, I want to put it on top, you know, like a film, this is food or that I take such a plan so here you are
I smear it like this, I do it like, but the gas is also not good, I just inform the layer what it is like at the junction, and this is how we finish looking for it, it went like this, the shura is glued like this, like this, like this, it was like the composition of the court, everything is like this
so it dried up for me for a day I did not pass it for about fifteen hours passed as if I smeared it three times here you see and we seem to hold on stopped to inflate this check or not at all it will fall plow this structure does not
annoying such a grandmother can not glue on it and the horseradish is not cut at all so that it does not crack here, I hope just, but at the junction I take just these, well, they hold the right for a share and so I keep it like it doesn’t break off, I’ll try to sew up the middle of the carriage it is desirable to take
would, but I think under load it is necessary that my specifically sting so I will try to sew it up and try it already with this cover, in general, I have already collected completely everything works fine here like this
just half thought at first I’ll do how to tell the time, but you know whether it’s working or if it works.
what is it such a support only I don't even know what it is it is in Chinese such find that, but this second one I have not fixed yet because there is a gust somewhere else, but if I have a gust somewhere here
here in this pimp it will not be done like that three times, but maybe he will see one number like this, so that's where to steam it, I don't even know better, maybe it’s easier to rip it from this side was where it’s so-and-so in this rough it my
ripped it from this side so that this is how everything works shakes without problems at least morally it will be, but no, look this thing works glue the globe drives everything inside shakes, but we also believe perry, let's night while he could
make a year ago, I get a photo I bought you see so that dude in the social service works fumbles for him families talking about the increase sold this thing, in principle, could and could fix it and now I have two options, now I will make this option and
I will make another fallback option 1 that noise does not work yet our stock is a little sho I could immediately change it always when I shoot a video I don’t adjust a little bit and change so everything I have is not normal even very 1779 which is 20 to 80
seventy-three blows like an astronaut like this guys so what a life hack it seems to me it can be any bicycle you can do it I don’t know, but with ordinary bicycle glue, this kind of red sun is not here
I know if he will roll, but he says take the globe, as if the moment is not a real moment to smell like a moment, absolutely everything is like a moment, but he glued me to it normally until time shows how much more this will last, of course, I had exactly the impulse
here at this place it turns out here like a fold is formed, although it spends all the time tugging, when you take pressure along with it, I take it like this and raise it, I thought that I had a couple somewhere here, even though they are here, this place is very happy even so
made reliably in india for this chinese one in that place I will still have a chinese booth to repair it, maybe I will shoot a video, but now, as usual, note that you have been bye bye be
How and what to glue a chamber for a tonometer
linar bikzhanov: so much empty chatter!
isli kamilova: I have a cuff with two horns, where can I find such a bag? I bought it for two days and that's it.
Georgi Georgi: In short, I fixed it, and went to buy a new one!
Vitya Borispol: Take the dropper bag; or to collect blood. They fit the cuff perfectly. I connected the pipe and sewed it up and plows normally!
Press CTRL + D to add the page to the bookmarks.
Press CTRL + D to bookmark the page.
How to glue the cuff (rubber) of the hatch in the washing machine?
One of the subscribers recommended a sealant to glue the door rubber in the washing machine. So I spent it.
Repairing the MPT automatik 90 tonometer does not inflate the cuffs.
Repair of the cuff by gluing.
Repair of the tonometer, replacement of the pear and hoses
Repairing a manual tonometer, as it turned out, is not so difficult if there are Chinese spare parts on sale from Aliexpress.
Replacing the camera in the cuff of the tonometer.
Replacing the camera in the cuff of the shoulder tonometer.
Budget repair of the tonometer cuff.
DIY tonometer cuff repair
How and how to glue a chamber for a tonometer.
repair of the cuff of an automatic tonometer
repair of the cuff of an automatic tonometer.
Tonometer. The pressure does not rise when pumped or drops rapidly when measured without a skip. Checked the connections.
This video was created for informational purposes only. From practice I add that the cuffs are in the tonometer.
Ecouter DIY tonometer cuff repair, telecharger DIY tonometer cuff repair, Music DIY tonometer cuff repair, DIY tonometer cuff repair Gratuit, chanson de DIY tonometer cuff repair, album DIY tonometer cuff repair.
الإستماع و تحميل اغنية DIY tonometer cuff repair, تنزيل موسيقى DIY tonometer cuff repair, اسمع طرب DIY tonometer cuff repair
Listen to DIY tonometer cuff repair, or Download MP3 song of DIY tonometer cuff repair, Artist Free Music track of DIY tonometer cuff repair
How to quickly reduce blood pressure at home - with and without drugs., Tanometer cuff repair, Life is great! Why the tonometer is wrong. (09/04/2017), DIY. Do-it-yourself cyclone filter (cyclone) in a home workshop., Making cabinet furniture. A couple of tricks., VLOG Dollar at 28 Life became more beautiful than SSD Samsung EVO Crosses from Second, Treatment with soda - Entries by Helena Roerich, why do you need to take soda, How to Measure Pressure without a Tonometer! A method that will be useful to everyone !, Sole burst? A professional approach to repair. Part-1, How to make 3D self-leveling floors. Full-length video. How to connect a dishwasher, Repairing heating filaments The easy way, Water heater with digital display, electric faucet Package., Natural pill for old age! Here's what you need to know. repair of the cuff of an automatic tonometer, How to choose a tonometer. Which tonometer is better to buy. How to disassemble the drill chuck. Drill Chuck Disassembly, Kitchen Making. from A to Z. part 1, Connecting the drain of the washing machine to the siphon of the washbasin, Installation of electrical wiring PART 1 | Electrical wiring in the apartment
Where to repair it, how and why to glue the tonometer cushion, how to properly glue the tonometer's rubber cushion, you will learn all this in this interesting video!
The video from my former channel, the channel has deleted, I think the video is useful, there were almost 20,000 views!
==========
Links to Cashback services:
MegaBonus + expansion
+ 250 rubles + extension
biggest% of AD + expansion
Cashback from Letyshops up to 40%
==========
My investments:
Hashflare promo code: CFBA1BC4-1
Genesis promo code: qGSHML
Hashing24
Maya Group
Business Forum
CRYP TRADE:
==========
Links to useful services:
Affiliates 24
Monitoring
==========
Additional services:
Workzilla -
Like Rock -
Kaleostra -
SocialPublic -
WMail -
SeoFast -
==========
Playlists online help:
Mining -
Programs -
Video (click to play). |
If necessary, I will provide online help, I will answer everyone!
Subscribe, like and write comments!
Sincerely,
Igor Pichugin