In detail: do-it-yourself repair of ut33c multimeters diagrams from a real master for the site my.housecope.com.
When repairing electronics, you have to carry out a large number of measurements with various digital instruments. This is an oscilloscope, an ESR meter, and what is used most often and without the use of which no repair can do: of course, a digital multimeter. But sometimes it happens that help is already required by the instruments themselves, and this happens not so much from the inexperience, haste or carelessness of the master, but from an annoying accident, such as happened to me recently.
DT Series Multimeter - Appearance
It was like this: after replacing the broken field-effect transistor during the repair of the LCD TV power supply, the TV did not work. An idea arose, which, however, should have come even earlier, at the diagnostic stage, but in a hurry it was not possible to check the PWM controller even for low resistance or a short circuit between the legs. It took a long time to remove the board, the microcircuit was in our DIP-8 package and it was not difficult to ring its feet on the short circuit even on top of the board.
Electrolytic capacitor 400 volt
I disconnect the TV from the mains, wait for the standard 3 minutes to discharge the capacitors in the filter, those very large barrels, electrolytic capacitors for 200-400 Volts, which everyone saw when disassembling a switching power supply.
I touch the probes of the multimeter in the mode of sound continuity of the PWM controller legs - suddenly a beep sounds, I remove the probes in order to call the rest of the legs, the signal sounds for another 2 seconds. Well, I think that's all: again 2 resistors burned out, one in the resistance measurement circuit of the 2 kOhm mode, for 900 Ohm, the second for 1.5 - 2 kOhm, which is most likely in the ADC protection circuits. I had already encountered a similar nuisance, in the past a friend hit me with a tester in the same way, so I did not get upset - I went to the radio store for two resistors in SMD cases 0805 and 0603, one ruble a piece, and soldered them.
Video (click to play).
Searches for information on the repair of multimeters on various resources, at one time, gave out several typical schemes, on the basis of which most models of cheap multimeters are built. The problem was that the reference designations on the boards did not match the designations on the found diagrams.
Burnt resistors on the multimeter board
But I was lucky, on one of the forums a person described in detail a similar situation, the failure of the multimeter when measuring with the presence of voltage in the circuit, in the mode of sound dialing. If there were no problems with the 900 Ohm resistor, several resistors on the board were connected in a chain and it was easy to find it. Moreover, for some reason it did not turn black, as is usually the case during combustion, and it was possible to read the denomination and try to measure its resistance. Since the multimeter contains precise resistors that have 4 digits in their designation, it is better, if possible, to change the resistors to exactly the same ones.
There were no precision resistors in our radio store and I took the usual one for 910 ohms. As practice has shown, the error with such a replacement will be quite insignificant, because the difference between these resistors, 900 and 910 Ohms, is only 1%. Determining the value of the second resistor was more difficult - from its terminals there were tracks to two transition contacts, with metallization, to the back of the board, to the switch.
Place for soldering thermistor
But I was lucky again: two holes were left on the board connected by tracks in parallel with the resistor leads and they were signed by RTS1, then everything was clear. The thermistor (РТС1), as we know from the pulse power supplies, is soldered in order to limit the currents through the diodes of the diode bridge when the pulsed power supply is turned on.
Since electrolytic capacitors, those very large barrels of 200-400 volts, at the moment the power supply is turned on and the first fractions of a second at the start of charging, behave almost like a short circuit - this causes large currents through the bridge diodes, as a result of which the bridge can burn out.
To put it simply, a thermistor has a low resistance in normal mode when small currents flow, corresponding to the mode of operation of the device. With a sharp multiple increase in current, the resistance of the thermistor also increases sharply, which, according to Ohm's law, as we know, causes a decrease in the current in the circuit section.
Resistor 2 Kom Ohm on the diagram
When repairing on the circuit, presumably we change to a 1.5 kΩ resistor, the resistor indicated on the circuit with a nominal value of 2 kΩ, as they wrote on the resource from which they took the information, during the first repair, its value is not critical and it was recommended to put it, nevertheless, at 1.5 kΩ.
We continue... After the capacitors are charged and the current in the circuit has decreased, the thermistor decreases its resistance and the device operates normally.
900 ohm resistor on the diagram
Why is a thermistor installed instead of this resistor in expensive multimeters? With the same purpose as in switching power supplies - to reduce large currents that can lead to the burnout of the ADC, arising in our case as a result of an error of the master conducting the measurements, and thereby protecting the analog-to-digital converter of the device.
Or, in other words, that very black drop, after the combustion of which the device usually no longer makes sense to restore, because this is a laborious task and the cost of parts will exceed at least half the cost of a new multimeter.
How can we solder these resistors - perhaps beginners who have not previously dealt with SMD radio components will think. After all, they most likely do not have a soldering hair dryer in their home workshop. There are three ways here:
First, you will need an EPSN soldering iron with a power of 25 watts, with a blade blade with a cut in the middle, in order to heat both terminals at once.
The second way, by biting off with side cutters, a drop of Rose or Wood's alloy, immediately on both contacts of the resistor, and flatten both of these terminals with a sting.
And the third way, when we have nothing but a 40-watt soldering iron of the EPSN type and the usual POS-61 solder - we apply it to both leads so that the solders mix and as a result, the total melting temperature of the lead-free solder decreases, and we heat both leads of the resistor alternately, while trying to move it a little.
Usually this is enough for our resistor to be sealed off and stick to the tip. Of course, do not forget to apply the flux, it is better, of course, liquid Alcohol rosin flux (GFR).
In any case, no matter how you dismantle this resistor from the board, bumps of old solder will remain on the board, we need to remove it using a dismantling braid, dipping it in an alcohol-rosin flux. We put the tip of the braid directly on the solder and press it, warming it up with the tip of the soldering iron until all the solder from the contacts is absorbed into the braid.
Well, then it's a matter of technology: we take the resistor we bought from the radio store, put it on the contact pads that we freed from the solder, press it down with a screwdriver from above and touch the pads and leads located on the edges of the resistor with the tip of a 25-watt soldering iron, solder it in place.
Solder Braid - Applications
The first time, it will probably turn out crooked, but the most important thing is that the device will be restored. On the forums, opinions about such repairs were divided, some argued that due to the cheapness of multimeters, it makes no sense to repair them at all, they say they threw it out and went to buy a new one, others were even ready to go all the way and re-solder the ADC). But as this case shows, sometimes repairing a multimeter is quite simple and cost-effective, and any home craftsman can easily handle such a repair. Successful repairs to everyone! AKV.
Like any other item, the multimeter can fail during operation or have an initial, factory defect that was not noticed during production. In order to find out how to repair a multimeter, you should first understand the nature of the damage.
Experts advise starting the search for the cause of the malfunction with a thorough examination of the printed circuit board, since short circuits and poor soldering are possible, as well as a defect in the leads of the elements along the edges of the board.
Factory defect in these devices is manifested mainly on the display. There can be up to ten types of them (see table). Therefore, it is better to repair digital multimeters using the instructions that come with the device.
The same breakdowns can occur after operation. The above malfunctions may also appear during operation. However, if the device operates in constant voltage measurement mode, it rarely breaks.
The reason for this is its overload protection. Also, the repair of a faulty device should begin with checking the supply voltage and the operability of the ADC: the stabilization voltage is 3 V and there is no breakdown between the power pins and the common ADC output.
Experienced users and professionals have repeatedly stated that one of the most likely causes of frequent breakdowns in the device is poor quality production. Namely, soldering contacts with acid. As a result, the contacts are simply oxidized.
However, if you are not sure what kind of breakdown caused the inoperative state of the device, you should still contact a specialist for advice or help.
It is impossible to imagine a repairman's workbench without a handy, inexpensive digital multimeter.
This article describes the device of the 830 series digital multimeters, its circuit, as well as the most common malfunctions and how to fix them.
Currently, a huge variety of digital measuring instruments of varying degrees of complexity, reliability and quality are produced. The basis of all modern digital multimeters is an integrated analog-to-digital voltage converter (ADC). One of the first such ADCs suitable for constructing inexpensive portable measuring instruments was a converter based on the ICL7106 microcircuit manufactured by MAXIM. As a result, several successful low-cost models of 830 series digital multimeters have been developed, such as M830B, M830, M832, M838. DT can be used instead of the letter M. This instrument series is currently the most widespread and most repeatable in the world. Its basic capabilities: measuring direct and alternating voltages up to 1000 V (input resistance 1 MΩ), measuring direct currents up to 10 A, measuring resistances up to 2 MΩ, testing diodes and transistors. In addition, in some models there is a mode of sound continuity of connections, temperature measurement with and without a thermocouple, generation of a meander with a frequency of 50 ... 60 Hz or 1 kHz. The main manufacturer of this series of multimeters is Precision Mastech Enterprises (Hong Kong).
The basis of the multimeter is the ADC IC1 of the 7106 type (the closest domestic analogue is the 572PV5 microcircuit). Its structural diagram is shown in Fig. 1, and the pinout for the version in the DIP-40 package is shown in Fig. 2. The 7106 core can be preceded by different prefixes depending on the manufacturer: ICL7106, ТС7106, etc. Recently, more and more often used are chipless microcircuits (DIE chips), the crystal of which is soldered directly to the printed circuit board.
Consider the circuit of the Mastech M832 multimeter (Fig. 3). Pin 1 of IC1 supplies a positive 9V battery supply voltage, and Pin 26 supplies a negative battery supply. Inside the ADC there is a 3 V stabilized voltage source, its input is connected to pin 1 of IC1, and the output is connected to pin 32. Pin 32 is connected to the common pin of the multimeter and is galvanically connected to the COM input of the device.The voltage difference between pins 1 and 32 is approximately 3 V in a wide range of supply voltages - from nominal to 6.5 V. This stabilized voltage is fed to the adjustable divider R11, VR1, R13, and from its output to the input of the microcircuit 36 (in the mode measurements of currents and voltages). The divider sets the potential U at pin 36, equal to 100 mV. Resistors R12, R25 and R26 perform protective functions. Transistor Q102 and resistors R109, R110 and R111 are responsible for indicating the discharge of the battery. Capacitors C7, C8 and resistors R19, R20 are responsible for displaying the decimal points of the display.
Operating input voltage range Umax directly depends on the level of the regulated reference voltage at pins 36 and 35 and is
The stability and accuracy of the display is dependent on the stability of this reference voltage.
The display N readings depend on the input voltage U and are expressed as a number
A simplified circuit of the multimeter in the voltage measurement mode is shown in Fig. 4.
When measuring DC voltage, the input signal is fed to R1… R6, from the output of which, through a switch [according to the scheme 1-8 / 1… 1-8 / 2), it is fed to the protective resistor R17. This resistor also forms a low-pass filter when measuring AC voltage together with the capacitor C3. Then the signal goes to the direct input of the ADC microcircuit, pin 31. The potential of the common pin, generated by the 3 V stabilized voltage source, pin 32, is fed to the inverse input of the microcircuit.
When measuring AC voltage, it is rectified by a half-wave rectifier on diode D1. Resistors R1 and R2 are selected so that when measuring sinusoidal voltage, the device shows the correct value. ADC protection is provided by the divider R1 ... R6 and the resistor R17.
A simplified circuit of the multimeter in the current measurement mode is shown in Fig. 5.
In the mode of measuring direct current, the latter flows through the resistors R0, R8, R7 and R6, which are switched depending on the measuring range. The voltage drop across these resistors through R17 is fed to the ADC input, and the result is displayed. ADC protection is provided by diodes D2, D3 (in some models they may not be installed) and fuse F.
A simplified circuit of the multimeter in the resistance measurement mode is shown in Fig. 6. In the resistance measurement mode, the dependence expressed by the formula (2) is used.
The diagram shows that the same current from the voltage source + U flows through the reference resistor and the measured resistor R "(the currents of inputs 35, 36, 30 and 31 are negligible) and the ratio of U and U is equal to the ratio of the resistances of the resistors R" and R ^. R1..R6 are used as reference resistors, R10 and R103 are used as current setting resistors. Protection of the ADC is provided by thermistor R18 (some cheap models use conventional 1.2 kΩ resistors), transistor Q1 in zener diode mode (not always installed) and resistors R35, R16 and R17 at inputs 36, 35 and 31 of the ADC.
Continuity mode The dialing circuit uses IC2 (LM358), which contains two operational amplifiers. A sound generator is assembled on one amplifier, and a comparator on the other. When the voltage at the input of the comparator (pin 6) is less than the threshold, a low voltage is set at its output (pin 7), which opens the switch on the transistor Q101, as a result of which a sound signal is emitted. The threshold is determined by the divider R103, R104. Protection is provided by resistor R106 at the comparator input.
All malfunctions can be divided into factory defects (and this happens) and damage caused by erroneous actions of the operator.
Since multimeters use tight wiring, shorts of elements, poor soldering and breakage of the leads of elements, especially those located at the edges of the board, are possible. Repair of a faulty device should begin with a visual inspection of the printed circuit board.The most common factory defects of M832 multimeters are shown in the table.
The LCD display can be checked for proper operation using a 50.60 Hz AC voltage source with an amplitude of several volts. As such a source of alternating voltage, you can take the M832 multimeter, which has a meander generation mode. To check the display, put it on a flat surface with the display up, connect one probe of the M832 multimeter to the common terminal of the indicator (bottom row, left terminal), and apply the other probe of the multimeter alternately to the rest of the display. If it is possible to get the ignition of all segments of the display, then it is serviceable.
The above malfunctions may also appear during operation. It should be noted that in the DC voltage measurement mode, the device rarely fails, because well protected from input overloads. The main problems arise when measuring current or resistance.
Repair of a faulty device should begin with checking the supply voltage and the ADC's operability: the stabilization voltage is 3 V and there is no breakdown between the power pins and the common ADC output.
In the current measurement mode when using the V, Q and mA inputs, despite the presence of a fuse, there may be cases when the fuse blows out later than the safety diodes D2 or D3 have time to break through. If a fuse is installed in the multimeter that does not meet the requirements of the instructions, then in this case the resistances R5 ... R8 may burn out, and this may not appear visually on the resistances. In the first case, when only the diode breaks through, the defect appears only in the current measurement mode: the current flows through the device, but the display shows zeros. In case of burnout of resistors R5 or R6 in the voltage measurement mode, the device will overestimate the readings or show an overload. When one or both resistors are completely burned out, the device is not reset in voltage measurement mode, but when the inputs are closed, the display is set to zero. When the resistors R7 or R8 burn out on the current measuring ranges of 20 mA and 200 mA, the device will show an overload, and in the 10 A range - only zeros.
In resistance measurement mode, faults usually occur in the 200 ohm and 2000 ohm ranges. In this case, when the voltage is applied to the input, the resistors R5, R6, R10, R18, the transistor Q1 can burn out and the capacitor C6 can break through. If the transistor Q1 is completely broken, then when measuring the resistance, the device will show zeros. In case of incomplete breakdown of the transistor, the multimeter with open probes will show the resistance of this transistor. In the modes of measuring voltage and current, the transistor is short-circuited by a switch and does not affect the readings of the multimeter. With a breakdown of capacitor C6, the multimeter will not measure voltage in the ranges of 20 V, 200 V and 1000 V or significantly underestimate the readings in these ranges.
If there is no indication on the display, when there is power to the ADC, or there is a visually noticeable burnout of a large number of circuit elements, there is a high probability of damage to the ADC. The ADC's serviceability is checked by monitoring the voltage of the 3 V stabilized voltage source. In practice, the ADC burns out only when a high voltage is applied to the input, much higher than 220 V. Very often, cracks appear in the compound of the open-frame ADC, the current consumption of the microcircuit increases, which leads to its noticeable heating ...
When a very high voltage is applied to the input of the device in the voltage measurement mode, a breakdown may occur in the elements (resistors) and on the printed circuit board, in the case of the voltage measurement mode, the circuit is protected by a divider on the resistances R1.R6.
For cheap models of the DT series, long part leads can be shorted to the screen located on the back cover of the device, disrupting the operation of the circuit. Mastech does not have such defects.
A source of a stabilized voltage of 3 V in an ADC for cheap Chinese models can in practice give a voltage of 2.6-3.4 V, and for some devices it stops working already at a voltage of a supply battery of 8.5 V.
The DT models use low quality ADCs and are very sensitive to the C4 and R14 integrator chain ratings. High-quality ADCs in Mastech multimeters allow the use of elements of close denominations.
Often, in DT multimeters, when the probes are open in the resistance measurement mode, the device approaches the overload value for a very long time ("1" on the display) or is not set at all. It is possible to "cure" a poor-quality ADC microcircuit by reducing the value of the resistance R14 from 300 to 100 kOhm.
When measuring resistances in the upper part of the range, the device "flushes" the readings, for example, when measuring a resistor with a resistance of 19.8 kOhm, it shows 19.3 kOhm. It is "treated" by replacing the capacitor C4 with a capacitor of 0.22 ... 0.27 μF.
Since cheap Chinese firms use low-quality unpackaged ADCs, there are frequent cases of broken pins, and it is very difficult to determine the cause of the malfunction and it can manifest itself in different ways, depending on the broken pin. For example, one of the indicator leads is off. Since multimeters use displays with static indication, then to determine the cause of the malfunction, it is necessary to check the voltage at the corresponding pin of the ADC microcircuit, it should be about 0.5 V relative to the common pin. If it is zero, then the ADC is faulty.
There are malfunctions associated with poor-quality contacts on the biscuit switch, the device works only when the biscuit is pressed. Firms that make cheap multimeters rarely coat the tracks under the rocker switch with grease, which is why they quickly oxidize. Often the tracks are dirty. It is repaired as follows: the printed circuit board is removed from the case, and the switch tracks are wiped with alcohol. Then a thin layer of technical petroleum jelly is applied. Everything, the device is repaired.
With DT series devices, it sometimes happens that the alternating voltage is measured with a minus sign. This indicates an incorrect installation of D1, usually due to incorrect marking on the diode body.
It happens that manufacturers of cheap multimeters put low-quality operational amplifiers in the sound generator circuit, and then when the device is turned on, a buzzing buzzer is heard. This defect is eliminated by soldering a 5 μF electrolytic capacitor parallel to the power supply circuit. If this does not ensure the stable operation of the sound generator, then it is necessary to replace the operational amplifier with the LM358P.
Often there is such a nuisance as battery leakage. Small drops of electrolyte can be wiped off with alcohol, but if the board is heavily flooded, then good results can be obtained by washing it with hot water and laundry soap. After removing the indicator and unsoldering the buzzer, using a brush, for example, a toothbrush, you need to thoroughly soap the board on both sides and rinse it under running water from the tap. After repeating the wash 2.3 times, the board is dried and installed in the case.
Most of the devices produced recently use DIE chips ADCs. The crystal is installed directly on the PCB and is filled with resin. Unfortunately, this significantly reduces the maintainability of the devices, because when the ADC fails, which is quite common, it is difficult to replace it. Unpackaged ADCs are sometimes sensitive to bright light. For example, if you work near a table lamp, the measurement error may increase. The fact is that the indicator and the board of the device have some transparency, and light, penetrating through them, enters the ADC crystal, causing a photoelectric effect. To eliminate this drawback, you need to remove the board and, after removing the indicator, glue the location of the ADC crystal (it is clearly visible through the board) with thick paper.
When buying DT multimeters, you should pay attention to the quality of the switch mechanics; be sure to rotate the multimeter's rocker switch several times to make sure that the switching occurs clearly and without jamming: plastic defects cannot be repaired.
Sergey Bobin. "Repair of electronic equipment" No. 1, 2003
Or sign in with these services
Your post must be verified by a moderator
It is quite within the power of every user who is well acquainted with the basics of electronics and electrical engineering to independently organize and repair the multimeter. But before embarking on such a repair, you must try to figure out the nature of the damage that has occurred.
It is most convenient to check the serviceability of the device at the initial stage of repair by inspecting its electronic circuit. For this case, the following troubleshooting rules have been developed:
it is necessary to carefully examine the printed circuit board of the multimeter, on which there may be clearly distinguishable factory defects and errors;
special attention should be paid to the presence of unwanted short circuits and poor-quality soldering, as well as defects on the terminals at the edges of the board (in the area of the display connection). For repairs, you will have to use soldering;
factory errors most often manifest themselves in the fact that the multimeter does not show what it should according to the instructions, and therefore its display is examined first of all.
If the multimeter gives incorrect readings in all modes and IC1 heats up, then you need to inspect the connectors to check the transistors. If the long leads are closed, then the repair will consist only in opening them.
In total, a sufficient number of visually detectable faults can accumulate. You can familiarize yourself with some of them in the table and then eliminate it yourself. (to the address: Before repairing, it is necessary to study the multimeter circuits, which are usually given in the passport.
If they want to check the serviceability and repair the multimeter indicator, then they usually resort to using an additional device that produces a signal of a suitable frequency and amplitude (50-60 Hz and units of volts). In its absence, you can use a multimeter type M832 with the function of generating rectangular pulses (meander).
To diagnose and repair the multimeter display, it is necessary to remove the working board from the device case and select a position convenient for checking the indicator contacts (screen up). After that, you should connect the end of one probe to the common terminal of the indicator under investigation (it is located in the bottom row, the far left), and alternately touch the other end to the signal outputs of the display. In this case, all of its segments should light up one after the other according to the wiring of the signal buses, which should be read separately. Normal "operation" of the tested segments in all modes indicates that the display is working properly.
Additional Information. This malfunction most often manifests itself during the operation of a digital multimeter, in which its measuring part fails and needs to be repaired extremely rarely (provided that the instructions are followed).
The last remark concerns only constant values, when measuring which the multimeter is well protected against overloads. Serious difficulties in identifying the reasons for the failure of the device are most often encountered when determining the resistances of the circuit section and in the dialing mode.
In this mode, typical malfunctions, as a rule, appear in the measuring ranges up to 200 and up to 2000 Ohm. When an extraneous voltage enters the input, as a rule, resistors under the designations R5, R6, R10, R18 burn out, as well as the transistor Q1. In addition, the capacitor C6 often breaks through. The consequences of exposure to extraneous potential are manifested as follows:
when the Q1 triode is completely "burnt out", when determining the resistance, the multimeter shows one zeros;
in case of incomplete breakdown of the transistor, the device with open ends should show the resistance of its junction.
Note! In other measurement modes, this transistor is short-circuited and therefore has no effect on the display.
With a breakdown of C6, the multimeter will not work at the measuring limits of 20, 200 and 1000 Volts (the option of a strong understatement of the reading is not excluded).
If the multimeter constantly beeps when dialing or is silent, then the reason may be poor-quality soldering of the pins of IC2. The repair consists in careful soldering.
Inspection and repair of an inoperative multimeter, the malfunction of which is not related to the cases already considered, is recommended to start with checking the voltage of 3 Volts on the ADC supply bus. In this case, first of all, it is necessary to make sure that there is no breakdown between the supply terminal and the common terminal of the converter.
The disappearance of indication elements on the display screen in the presence of a supply voltage converter with a high degree of probability indicates damage to its circuit. The same conclusion can be made when a significant number of circuit elements located near the ADC are burned out.
Important! In practice, this node "burns out" only when a sufficiently high voltage (more than 220 volts) hits its input, which manifests itself visually in the form of cracks in the module compound.
Before talking about repairs, you need to check. A simple way to test the ADC for suitability for further operation is to dial its terminals using a known working multimeter of the same class. Note that the case when the second multimeter incorrectly shows the measurement results is not suitable for such a check.
When preparing for operation, the device is switched to the diode “ringing” mode, and the measuring end of the wire in red insulation is connected to the “minus power” output of the microcircuit. Following this black probe, each of its signal legs is sequentially touched. Since there are protective diodes at the inputs of the circuit, connected in the opposite direction, after applying a forward voltage from a third-party multimeter, they should open.
The fact of their opening is recorded on the display in the form of a voltage drop across the junction of the semiconductor element. Similarly, the circuit is checked when a probe in black insulation is connected to pin 1 (+ ADC power supply) and then touching all other pins. In this case, the indications on the display screen should be the same as in the first case.
When the polarity of the connection of the second measuring device is changed, its indicator always shows an open circuit, since the input resistance of the working microcircuit is large enough. In this case, the conclusions will be considered faulty, in both cases showing the final resistance value. If, for any of the described connection options, the multimeter shows an open circuit, this most likely indicates an internal circuit break.
Since modern ADCs are most often produced in an integral version (without a case), it is rare for anyone to replace them. So if the converter is burnt out, then the multimeter cannot be repaired, it cannot be repaired.
Repair will be required if there are malfunctions associated with the loss of contact in the rotary switch. This is manifested not only in the fact that the multimeter does not turn on, but also in the impossibility of obtaining a normal connection without pressing hard on the biscuit. This is explained by the fact that in cheap Chinese multimeters, the contact tracks are rarely covered with high-quality grease, which leads to their rapid oxidation.
When used in dusty conditions, for example, they become dirty over time and lose contact with the switch strip. To repair this multimeter unit, it is enough to remove the printed circuit board from its case and wipe the contact tracks with a cotton swab dipped in alcohol. Then a thin layer of high-quality technical petroleum jelly should be applied to them.
In conclusion, we note that if factory "missing" or contact closures are detected in the multimeter, these shortcomings should be eliminated using a low-voltage soldering iron with a well-honed tip. If you are not completely sure of the reason for the breakdown of the device, you should contact a repair specialist for measuring equipment.
He took off the protective cover, spun it up - and was stunned when he saw the board! As I understand it, some of the connections, including the power switch, were soldered by the Chinese using soldering acid (the one etched on zinc).
Naturally, the switch is covered in white and the contacts have oxidized. I wiped the leads with alcohol, took the rosin and warmed up the switch leads. Now the conclusions gleamed with a tinny natural color!
On the reverse side, there are a bunch of microcircuits and parts, the display is held on a flexible cable that can be unscrewed if necessary. Mine is already patsarab, because for more than one year I was dragging along with the instrument in the box. For carrying in the future, I will use my own box of a digital multimeter, so as not to scratch the device even more.
Let's go back to the repair. All conclusions were heated, checked - it works! We assemble carefully in the reverse order, we do not press the screws too much - a fragile board may crack (then the tracks will have to be soldered later).
We twist the case, tighten the cover and check the operability and accuracy, for example, by measuring the voltage of the old battery of the crown, mine turned out to be hooked.
Successful repairs to all! Comrade was with you. vanesex
Radio amateurs periodically encounter the problem of a multimeter breakdown. Most often, the problem is that the multimeter was soldered using acid and the contacts are simply oxidized. In this case, it is very easy to fix the problem, but there is a more serious problem, for example (as in my case), forgetting to discharge the capacitor, they put it in a digital multimeter and want to measure the capacity, after which the tester refuses to measure anything at all.
Having opened the multimeter, we obviously will not see anything, since the microcircuit was killed by static. The microcircuit itself will most likely be 324, as in the photo. Principled DT9205A schematic can be downloaded here.
But since the multimeter is made in China, then most likely we will not find any data on this microcircuit. So at first I did not find anything, but then I decided to search, having entered not all the elements of the microcircuit inscription, but only numbers. And the result made me happy - the microcircuit turned out to be lm324, or rather a Chinese copy, only with different letters. It is possible to change it to some other op-amp. If you have a radio store in the city, then you can quickly go there and buy this microcircuit, but if there is no such store (as in my case) or it is far away, and the capacitance meter is very necessary, then we change it to any existing microcircuit that contains in itself 4 operational amplifiers. If there are no quadruples, just put two microcircuits that contain 2 op amps, as I did at first.
True, it later turned out that with them the multimeter gives an error. This was due to the fact that the gain of my op amps was different from the gain of lm324. But there was nowhere to go, since I already said earlier we do not have radio stores, and ordering on the Internet is also not the best option - it will take a long time for the order to arrive, and I decided to put others. Just a couple of days before the repair of the DT9205A multimeter, an order of five TL074s arrived.
True, I had them in a DIP case and so that it did not interfere with the closing of the lid DT9205A - I soldered it with wires.
Perhaps when you change the op-amp, even if it is lm324, then the multimeter will show a mute not correctly. In this case, if the deviation is not very large, then this error is removed by a trimming resistor next to the microcircuit (shown by the red arrow), but since there may be deviations in the capacitor rating, it is better to measure its capacitance on another multimeter and adjust yours to the same reading.
And finally, a couple of pictures of the work after the renovation.
Enough time has passed since then - and the multimeter works without problems.I wish you all creative success! Article author: 13265
Analog multimeters were very quickly pushed out of the market by devices based on ADCs (analog-to-digital converters). This happened for a number of objective reasons (compact size, high accuracy, clarity of the provided result, acceptable cost, etc.), however, such measuring devices also have a number of disadvantages.
And the most significant is the complexity of the repair.
Firstly, modern manufacturers are very reluctant to share the schematic diagrams of devices, which greatly complicates the troubleshooting.
And, secondly, the microcircuit underlying the device is difficult not only to diagnose, but also to replace (often the crystal is not just soldered to the board, but also additionally filled with solid glue, which protects the crystal and also increases heat transfer).
Description of multimeters DT 832
The 830 series multimeters are very popular. They combine wide functionality and low cost. These devices are based on the ICL1706 ADC IC developed by MAXIM. Although at the moment there are many analogues from competitors, there is even a Russian implementation - 572PV5).
The original series of measuring instruments is marked as M832, the DT modification is a cheap analogue from Chinese manufacturers. Nevertheless, the functionality and the main scheme are preserved.
The multimeters are suitable for measuring voltages from 200 mV to 1 kV (for DC), currents from 200 μA to 10 A and resistances from 200 Ohm to 2 MΩ.
So, the main radioelements are indicated in the diagram below.
Rice. 1. Schematic diagram
To understand the basic logical connections between the nodes of the device, you can study the functional diagram.
Rice. 2. Functional diagram
It is best to take out the conclusions of the microcontroller separately.
The most interesting thing is that, even with a schematic diagram in hand, it will be very problematic to fix a multimeter. To understand why this is happening, it's easier to see everything once.
Rice. 4. Microcircuit underlying the device
The microcircuit is flooded, and the contacts are not indicated in any way, which significantly complicates the ringing of problematic elements, the control points are not indicated.
Due to the fact that there are a lot of reasons for breakdowns, below we will consider the most common.
Rice. 5. Fixing parts of the device
1. Broken switch... Due to the poor quality of the lubricant, literally after a few years, there may already be a noticeable difficulty in switching the mode. Another common problem is the fallout of the pressure balls (pictured above). In this case, the device stops working altogether, and a characteristic noise is heard in the case when shaking. The defect is repaired by simple reassembly and lubrication (it is best to use silicone) of the switch.
2. Burnout of individual elements... A very popular type of breakdown, when during the measurement process the switch is not moved to the desired position, and the resulting load exceeds the permissible value. In this case, in certain types of measurements, there are problems with the correctness of the received data. For diagnostics, you must have a circuit with known parameters or another working multimeter. When disassembling, it is very easy to find a burnt element. It will turn black. The problem is solved by replacing it with a full analogue (it is necessary to use the schematic diagram above to clarify the nominal value).
3. The screen goes out (when turned on, it lights up normally, but later it goes out smoothly)... The problem is most likely in the clock generator. In this case, the driving elements of the oscillatory circuit are C1 and R15. They must be checked and replaced if necessary.
4. The screen goes out, but with the cover removed it works as expected... With a high probability, the back cover touches the resistor R15 with the contact spring and short-circuits the master oscillator. The problem is solved by shortening the spring (or bending it).
5. In voltage measurement mode, the readings change spontaneously from 0 to 1... Most likely a problem with the integrator circuit. You can check and, if necessary, replace capacitors C2, C4, C5 and resistance R14.
6. In resistance measurement mode, the readings are set for a long time... Check and replace C5.
7. The data on the display is cleared for a long time... Most likely the problem is in the capacitor C3 (if the capacity is normal, it can be replaced with an analogue with a reduced absorption coefficient).
8. In any of the selected modes, the multimeter does not work correctly, the microcircuit itself heats up... It is necessary first of all to check if there is a short circuit in the terminals connected to the connector to test the transistors. You can look for a short circuit in other places in the circuit.
9. Individual segments disappear and appear on the LCD... With a high degree of probability, the conductivity through the rubber inserts (through which the display is connected to the board) has deteriorated. It is required to disassemble the connection, wipe the contacts with alcohol, tin the contact pads on the board if necessary.
This is not a complete list of possible malfunctions. A thorough visual inspection of the device, analysis of control point indicators and ringing of hotel elements will help to find them. For verification with the "norm" it is best to have on hand a known working DT 832 (as a reference).
Evgeniy / 09/14/2018 - 17:12 The schematic diagram does not correspond to either the photograph (or the model itself).
Alexander / 06/25/2018 - 13:59 multimeter DT832 board 8671 (832.4c-110426) the photo matches my multimeter, but in the diagram the resistors do not match the number of ohms. For example, I have 6R4 = 304, 6Rt1 = 102.6R3 = 105, 6R2 = 224, Rx2 = 205, and there are other numbers in the diagram above.